Wearable ECG patches
No source read defines a wearable ECG patch as a class. The nearest MeSH heading is Wearable Electronic Devices, which MeSH describes as electronic implements worn on the body. This page therefore describes one product only, a single-use ECG (electrocardiogram) monitor cleared in 2025, from the 510(k) summary its manufacturer wrote. That summary says the monitor records without a break for as long as 14 days, is put on and switched on by the patient, either at home or in a clinic, and is stuck to the upper left chest (the pectoral region). It stores the recording in its own memory, sends nothing wirelessly while worn, and goes back to the company for analysis afterwards; the summary says it is not meant for real-time monitoring. Its listed use is adults aged 18 or over, with or without passing symptoms, including palpitations, breathlessness, dizziness, lightheadedness, near-fainting, fainting, fatigue or anxiety. Other patches may differ. No detection rates are recorded.
Also called ECG patch · ECG patches · Wearable ECG patch
- Technology
- In this atlas
Overview
Technical summary
MeSH D000076251 (Wearable Electronic Devices), broad: its scope note lists wearable diagnostic devices among the examples and does not mention ECG (electrocardiogram). The only product read is K243650 (decision date 2025-08-15, product code DSH, which the US regulator names Recorder, Magnetic Tape, Medical; class 2, 21 CFR 870.2800). One cleared product does not define the class, so every device statement here is about that product and comes from the manufacturer's 510(k) summary.
How it works
In the one cleared product read (K243650), an adhesive monitor on the upper left chest records a single ECG (electrocardiogram) channel into its own memory for as long as 14 days. After the wear period the device is returned to the manufacturer. The recording is then analyzed by an analysis system, and qualified cardiac technicians at an independent diagnostic testing facility review it and create an end-of-wear report (the manufacturer's 510(k) summary).
Classification codes
-
D000076251 -
DSHRecorder, Magnetic Tape, Medical -
21 CFR 870.2800
Strengths
- The product read records without a break for as long as 14 days.
- The patient puts it on and starts it, either at home or in a clinic, and pressing a button on the device marks when a symptom occurs.
Limitations
- The manufacturer's summary says the product is not meant for real-time monitoring: it sends no data wirelessly during wear, and results come after the device is returned.
- The summary supplies indications and design only. It reports no detection or accuracy figures.
- Everything here rests on one cleared product, so other patches may differ in wear time, channels and review process.
Required infrastructure
- A prescription: the manufacturer's summary calls the product prescription-only. After wear, the device is returned, and technicians at a diagnostic testing facility review the analyzed recording and create a report.
Claims and evidence
No claims are published for this technology yet.
Record details
- Record id
cardio:wearable-ecg-patches- Content hash
38b0adbc4a881788ce516397e97e8223943121cf941a22ffb37a208b3b36d9c9