Cuffless blood pressure monitoring

Many devices that measure blood pressure without an inflating cuff are on sale, and they claim to be accurate. A European hypertension expert group says they could change how high blood pressure is diagnosed and managed, but that questions about their accuracy must be answered first. Its 2022 statement says recent European Society of Hypertension guidelines do not recommend them for diagnosis or management, and its 2023 statement sets out six tests for checking them. The British and Irish Hypertension Society says that, until it has a way to validate wearables, none can receive its endorsement or approval. This page records that position. It does not say that any such device is accurate.

Also called Cuffless wearable blood pressure devices

  • Idea
  • In this atlas

Overview

Technical summary

Hypothesis page with maturity set to early feasibility. The two European Society of Hypertension Working Group statements (2022 and 2023) were read as abstracts only, and neither names a patient population. The 2022 abstract calls cuffless monitors "a wide and heterogeneous group of novel technologies and devices with different intended uses". Some need individual cuff calibration, which the 2023 tests address through a recalibration test. The British and Irish Hypertension Society page says it has no formal validation process for wearables and that, until it has one, no wearable blood pressure device can receive its endorsement or approval. Newer studies of cuffless devices since 2023 were not read.

Hypothesis

Hypothesis, not a result. Nothing here has been demonstrated.

Cuffless blood pressure devices could support the diagnosis and management of high blood pressure once a validation standard shows that they are accurate.

Maturity
Early feasibility
Rationale
The European Society of Hypertension Working Group says these devices have "considerable potential to improve the awareness, treatment, and management of hypertension" and "considerable potential for changing the diagnosis and management of hypertension".
Proposed test
The 2023 European Society of Hypertension recommendations define six validation tests for intermittent cuffless devices: static, device position, treatment, awake and asleep, exercise and recalibration. Not every test is required for every device. A device would first need to pass the tests that apply to it.

Objections

  • The 2022 statement says the established validation protocols for cuff devices are inadequate for cuffless devices.
  • The 2022 statement says fundamental questions about accuracy, performance and implementation must be addressed before these devices can be recommended for clinical use.
  • The 2023 statement says there were no generally accepted validation protocols for them at that time.
  • The British and Irish Hypertension Society says that, until it has a way to validate wearables, none can receive its endorsement or approval.

Claims and evidence

Context People the European Society of Hypertension cuffless-device statements address (population not stated)

Recipient
Recipient not specified
Population
Not stated in the fetched abstracts. Neither abstract names a patient population.
Care setting
Not stated in the fetched abstracts
Geography
Not stated in the fetched abstracts
Life stage
Not stated in the fetched abstracts
Gestational age
Not reported
Postnatal age
Not reported
Follow-up
Not applicable (expert statements)
Eligibility: inclusion criteria not stated; exclusion criteria not stated
Included
None stated
Excluded
None stated
  • Description Guidance Certainty not assessed

    A 2023 statement by the European Society of Hypertension Working Group on Blood Pressure Monitoring and Cardiovascular Variability recommends six validation tests for intermittent cuffless blood pressure devices, which take readings at intervals longer than 30 seconds, usually every 30 to 60 minutes, or when the user starts a reading. The tests are static, device position, treatment, awake and asleep, exercise and recalibration. Not every test is required for every device. The statement says that, to date, there were no generally accepted validation protocols for these devices.

    Applies to Recipient not specified
    • Not stated in the fetched abstracts. Neither abstract names a patient population.
    • Not stated in the fetched abstracts
    • Not stated in the fetched abstracts
    • Not stated in the fetched abstracts
    • Gestational age Not reported
    • Postnatal age Not reported
    Follow-up Not applicable (expert statements)
    Certainty method
    Not assessed
    Limitations
    Abstract only: the full text was not fetched, so the pass criteria, sample sizes and which test applies to which device are not described here. The abstract names no patient population, so the population is not stated here. It covers intermittent cuffless devices only. The statement proposes how to validate devices. It is not a finding that any device has passed. Work after December 2023 was not checked.
    Last checked
    2026-10-11
    Next evidence check due
    Not scheduled
    Sources and citations (6)
    1. Supports European Society of Hypertension recommendations for the validation of cuffless blood pressure measuring devices (Stergiou and colleagues, Journal of Hypertension)

      Publication · Wolters Kluwer Health (Journal of Hypertension) · Edition Journal of Hypertension, December 2023 issue. The PubMed record and its abstract only. The full text was not fetched. · Published 2023-12-01 · pmid: 37303198

      Locator
      Abstract, sentence 5, first part
      How it applies
      PubMed abstract of the 2023 statement.
      Retrieved
      Source effective date
      Not reported
      Source state
      Current
      Snapshot hash
      0650ccca93af714940e8113b3609a54f02afabd9ecf98ab8970d83ebf23e3f8a
      Reuse basis
      The abstract carries a publisher copyright line, so only short quotations with attribution are used and only the sha256 of the PubMed record is kept. No file is committed.
    2. Supports European Society of Hypertension recommendations for the validation of cuffless blood pressure measuring devices (Stergiou and colleagues, Journal of Hypertension)

      Publication · Wolters Kluwer Health (Journal of Hypertension) · Edition Journal of Hypertension, December 2023 issue. The PubMed record and its abstract only. The full text was not fetched. · Published 2023-12-01 · pmid: 37303198

      Locator
      Abstract, sentence 5, second part
      How it applies
      PubMed abstract of the 2023 statement. First three tests.
      Retrieved
      Source effective date
      Not reported
      Source state
      Current
      Snapshot hash
      0650ccca93af714940e8113b3609a54f02afabd9ecf98ab8970d83ebf23e3f8a
      Reuse basis
      The abstract carries a publisher copyright line, so only short quotations with attribution are used and only the sha256 of the PubMed record is kept. No file is committed.
    3. Supports European Society of Hypertension recommendations for the validation of cuffless blood pressure measuring devices (Stergiou and colleagues, Journal of Hypertension)

      Publication · Wolters Kluwer Health (Journal of Hypertension) · Edition Journal of Hypertension, December 2023 issue. The PubMed record and its abstract only. The full text was not fetched. · Published 2023-12-01 · pmid: 37303198

      Locator
      Abstract, sentence 5, last part
      How it applies
      PubMed abstract of the 2023 statement. Last three tests.
      Retrieved
      Source effective date
      Not reported
      Source state
      Current
      Snapshot hash
      0650ccca93af714940e8113b3609a54f02afabd9ecf98ab8970d83ebf23e3f8a
      Reuse basis
      The abstract carries a publisher copyright line, so only short quotations with attribution are used and only the sha256 of the PubMed record is kept. No file is committed.
    4. Supports European Society of Hypertension recommendations for the validation of cuffless blood pressure measuring devices (Stergiou and colleagues, Journal of Hypertension)

      Publication · Wolters Kluwer Health (Journal of Hypertension) · Edition Journal of Hypertension, December 2023 issue. The PubMed record and its abstract only. The full text was not fetched. · Published 2023-12-01 · pmid: 37303198

      Locator
      Abstract, sentence 6
      How it applies
      PubMed abstract of the 2023 statement.
      Retrieved
      Source effective date
      Not reported
      Source state
      Current
      Snapshot hash
      0650ccca93af714940e8113b3609a54f02afabd9ecf98ab8970d83ebf23e3f8a
      Reuse basis
      The abstract carries a publisher copyright line, so only short quotations with attribution are used and only the sha256 of the PubMed record is kept. No file is committed.
    5. Supports European Society of Hypertension recommendations for the validation of cuffless blood pressure measuring devices (Stergiou and colleagues, Journal of Hypertension)

      Publication · Wolters Kluwer Health (Journal of Hypertension) · Edition Journal of Hypertension, December 2023 issue. The PubMed record and its abstract only. The full text was not fetched. · Published 2023-12-01 · pmid: 37303198

      Locator
      Abstract, sentence 3
      How it applies
      PubMed abstract of the 2023 statement.
      Retrieved
      Source effective date
      Not reported
      Source state
      Current
      Snapshot hash
      0650ccca93af714940e8113b3609a54f02afabd9ecf98ab8970d83ebf23e3f8a
      Reuse basis
      The abstract carries a publisher copyright line, so only short quotations with attribution are used and only the sha256 of the PubMed record is kept. No file is committed.
    6. Background European Society of Hypertension recommendations for the validation of cuffless blood pressure measuring devices (Stergiou and colleagues, Journal of Hypertension)

      Publication · Wolters Kluwer Health (Journal of Hypertension) · Edition Journal of Hypertension, December 2023 issue. The PubMed record and its abstract only. The full text was not fetched. · Published 2023-12-01 · pmid: 37303198

      Locator
      Abstract, sentence 4, definition of intermittent
      How it applies
      Defines which cuffless devices the tests are for.
      Retrieved
      Source effective date
      Not reported
      Source state
      Current
      Snapshot hash
      0650ccca93af714940e8113b3609a54f02afabd9ecf98ab8970d83ebf23e3f8a
      Reuse basis
      The abstract carries a publisher copyright line, so only short quotations with attribution are used and only the sha256 of the PubMed record is kept. No file is committed.

Record details

Record id
cardio:idea-cuffless-bp-monitoring
Content hash
82b83f229e307323fcc13abaa089e5bc9b68b736d0a98e589e37a0be69d6f87f